Digital Calibration Certificate

PTX-2021-000012

Recalibration Due

Sartorius Picus 8-Channel

Electronic air-displacement, multichannel

Serial number
SB-7790015
Channels
8
Volume range
5–100 µL
Registered lab
Vantage Cell Therapies
Location
Cambridge, MA
Overall result
Pass
Scan to verify
Performed
June 30, 2023
Next due
June 29, 2024
Technician
M. Alvarez, Lead Metrologist
Standard
ISO 8655-2 / ISO 8655-6
Method
Gravimetric method per ISO 8655-6, 10 replicates per test volume, all channels tested individually
Reference balance
Mettler Toledo XPR2U Micro Balance
Conditions
22.1°C · 51% RH · 1011 hPa

Technician notes: Initial calibration on enrollment.

ISO 8655 measurement results

Gravimetric test at 10%, 50%, and 100% of nominal volume, per ISO 8655-6.

Test pointMean measuredSystematic errorRandom error (CV)Result
10% – 10 µL10.22 µL+2.20% / ±4%2.10% / ≤2.50% Pass
50% – 50 µL50.40 µL+0.80% / ±1.80%0.95% / ≤1.30% Pass
100% – 100 µL100.65 µL+0.65% / ±1.50%0.65% / ≤1% Pass

Calibration history

Every past calibration for this instrument is its own permanent certificate — click any entry to see its full measurement data.

  1. July 4, 2025Calibrated — pass

    J. Okafor, Calibration Technician — Channel 4 O-ring replaced prior to test; all 8 channels retested to spec. Values below are the mean across channels.

    PTX-2023-000047

  2. June 29, 2024Calibrated — pass

    M. Alvarez, Lead Metrologist

    PTX-2022-000559

  3. June 30, 2023Calibrated — passViewing now

    M. Alvarez, Lead Metrologist — Initial calibration on enrollment.

    PTX-2021-000012